A novel acoustic uroflowmetry-based mobile app voiding diary: Comparison with conventional paper-based voiding diary

Kim J1, Lee S1

Research Type

Clinical

Abstract Category

E-Health

Abstract 145
On Demand E-Health
Scientific Open Discussion Session 16
On-Demand
Voiding Diary New Devices Benign Prostatic Hyperplasia (BPH)
1. Seoul National University Bundang Hospital
Presenter
J

Jung Kwon Kim

Links

Abstract

Hypothesis / aims of study
In our previous study, we introduced a smartphone-based uroflowmetry (UFM) device using acoustic analysis and reported a result comparable to that of contemporary office-based UFM. In the current study, we evaluated the usefulness of a novel acoustic UFM-based mobile app voiding diary (VD), which could record the voided volume automatically, by comparison with a conventional paper-based VD. We specifically focused on the (1) compliance and (2) correlation between the two methods in male patients with lower urinary tract symptoms and benign prostatic hyperplasia.
Study design, materials and methods
A total of 113 patients were included and screened between December 2019 and June 2020. Among them, 35 patients were excluded because they withdrew informed consent (n = 5), were females (n = 5), did not complete both VDs (n = 13), or had no app-supported smartphone (n = 12). Finally, 78 patients were enrolled, and a subsequent review of all data was performed. The analyzed data were as follows: (1) survey of convenience/satisfaction/preference comparing the two methods, (2) compliance regarding the completeness of both methods, and (3) the correlation of each metric (24-hour urine volume, nocturnal urine volume, nocturnal polyuria index, total number of voids, and the number of daytime voids, number of nocturnal voids, and maximal bladder capacity) between the two methods.
Results
The survey results of convenience, satisfaction, and preference were as follows. On a scale of 0 to 10, the ratings for all three questions were higher than 8 in the acoustic UFM-based mobile VD app, and the rating for the question on the overall preference of the method was as high as 9. We also found a good correlation between the two methods for nocturnal urine volume (r = 0.55, p = 0.04), nocturnal polyuria index (r = 0.66, p = 0.23), the total number of voids (r = 0.9, p = 0.02), number of nocturnal voids (r = 0.83, p = 0.02), and maximal bladder capacity (r = 0.89, p = 0.04) (Figure).
Interpretation of results
To the best of our knowledge, our acoustic UFM-based mobile VD app was the first portable device recording voided volume automatically. In our previous study, we introduced a smartphone-based UFM device using acoustic analysis and reported a result comparable to that of contemporary office-based UFM [1]. The results showed an excellent correlation between acoustic and standard UFM with regard to maximum and average flow rates as well as voided volumes. In the current study, we expanded the research on VDs, and again confirmed the accuracy of the acoustic method. The current novel acoustic UFM-based mobile VD app could also provide longitudinal trends in the urodynamic parameters in a quantitative manner, which will be useful for healthcare providers and payers who need to pre-screen and monitor LUTS and BPH patients.
Concluding message
The acoustic UFM-based mobile VD app demonstrated favorable findings compared to the conventional paper-based VD. Future large-scale prospective studies are needed to further validate our results.
Figure 1 Figure 1-1
Figure 2 Figure 1-2
References
  1. Lee YJ, Kim MM, Song SH, Lee S. A Novel Mobile Acoustic Uroflowmetry: Comparison With Contemporary Uroflowmetry. Int Neurourol J. 2021 Jan 4. doi: 10.5213/inj.2040250.125. Online ahead of print.
Disclosures
Funding This work was supported by the Korea Medical Device Development Fund grant funded by the Korea government (theMinistry of Science and ICT, the Ministry of Trade, Industry and Energy, theMinistry of Health & Welfare, the Ministry of Food and Drug Safety)(9991006814, KMDF_PR_20200901_0141) Clinical Trial No Subjects Human Ethics Committee Institutional Review Board (IRB) approval of Seoul National University Bundang Hospital (Approval number: B-1912/585-301) Helsinki Yes Informed Consent Yes