The artificial urinary sphincter (AUS) is the gold standard for treatment of post prostatectomy stress urinary incontinence (SUI). A subset of patients who have an AUS placed will subsequently require transurethral intervention with a resectoscope for treatment of a variety of conditions including bladder neck contracture (BNC) with or without associated dystrophic calcifications, bladder tumor, or intravesical clot burden. With improved survivorship after treatment of prostate cancer in the modern era, more patients with an indwelling AUS will likely require such procedures utilizing a resectoscope.
Passage of large instruments, such as standard resectoscopes (which are usually 24-28Fr in circumference) through the AUS cuff for endoscopic treatment is not possible for cuffs < 5 cm. Furthermore, even with successful navigation in cuffs > 5 cm, such instrumentation risks cuff trauma with subsequent cuff erosion, device damage, and/or infection.
Current options for resectoscope placement in such situations include a suprapubic approach using a percutaneously placed access sheath as well as “unbuckling” of the existing AUS cuff. Suprapubic access is invasive and potentially morbid due to risk of bowel injury and other complications. Alternatively, deactivating AUS and unbuckling the cuff through a perineal incision before instrumentation, followed by rebuckling the device postoperatively has been suggested. However, unbuckling the AUS cuff adds considerable time for the procedure, procedural complexity, as well as risk of wound complications and device infection [1].
As an alternative, we have repurposed and utilized a smaller diameter gynecological resectoscope to address this clinical problem. The “Princess”® is a 21Fr resectoscope (Richard Wolf Medical Instruments Corporation, Vernon Hills, IL, USA) primarily used for gynecologic intrauterine procedures (hysteroscopic myomectomy, polypectomy, endometrial ablation). It is compatible with bipolar cautery to minimize risk of thermal injury to AUS implant. It has continuous flow irrigation capability and a working channel that accommodates instruments up to 7Fr. We have utilized this device transurethrally for several cases successfully without the need for unbuckling of the AUS cuff.